19
Attaching a Cassette
WARNING:
• If you are using a CADD™ Administration Set or Medica-
tion Cassette Reservoir that does not have the flow stop
feature (reorder number does not start with 21-73xx): You
must use a CADD
™
Extension Set with Anti-Siphon Valve
or a CADD
™
Administration Set with either an integral or
add on Anti-Siphon Valve to protect against unregulated
gravity infusion that can result from an improperly at-
tached cassette. Unregulated gravity infusion can result in
death or serious injury.
• For detailed instructions and warnings pertaining to the
Medication Cassette Reservoir or CADD
™
Administration
Set, please refer to the instructions for use supplied with
those products.
• Frozen medication must be thawed at room temperature
only. Do not heat the Medication Cassette Reservoir in a
microwave oven as this may damage the medication, the
Medication Cassette Reservoir, or cause leakage.
Use aseptic technique as
instructed by your clinician.
1. Clamp the tubing on the
new Medication
Cassette Reservoir or
CADD
™
Administration
Set. If required, remove
blue clip from flow stop
feature.
2. Insert the cassette hooks into the hinge pins on the pump.