• If the pump is dropped or hit, the battery door may be-
come broken or damaged. Do not use the pump if the
battery door is damaged because the battery will not be
properly secured; this may result in loss of power or non-
delivery of drug.
• Prior to starting infusion, inspect the fluid path for kinks, a
closed clamp, or other upstream obstructions, and remove
all air to prevent air embolism.
• Close the tubing clamp before removing the cassette from
the pump to prevent unregulated gravity infusion.
• If you are using a CADD™ Administration Set or Medica-
tion Cassette Reservoir that does not have the flow stop
feature (reorder number does not start with 21-73xx): You
must use a CADD
™
Extension Set with Anti-Siphon Valve or
a CADD
™
Administration Set with either an integral or add
on Anti-Siphon Valve to protect against delivery inaccura-
cies and unregulated gravity infusion that can result from
an improperly attached cassette.
• For detailed instructions and warning pertaining to Medi-
cation Cassette Reservoirs or CADD
™
Administration Sets,
please refer to the instructions for use supplied with those
products.
• Frozen Medication must be thawed at room temperature
only. Do not heat the Medication Cassette Reservoir in a
microwave oven as this may damage the medication, the
Medication Cassette Reservoir, or cause leakage.
• Attach the cassette (the part of the Medication Cassette
Reservoir or CADD
™
Administration Set that attaches to
the pump) properly. An improperly attached or detached