9
has been damaged because the battery will not be prop-
erly secured; this may result in loss of power, nondelivery
of drug, and, depending on the type of drug being admin-
istered, death or serious injury.
To install a new battery:
1. Stop the pump by pressing
⁄
.
2. When you see
Stop the pump?
,
press
´
.
3. Press the button on the battery
door and slide the battery door
forward. Remove the used
battery.
4. Match the
+
and
–
markings on
the new battery with the mark-
ings on the pump. Insert the
battery. The pump will beep
if the battery is inserted
correctly.
5. Replace the battery door. The
pump will power up auto-
matically.
6. Start the pump by pressing
⁄
.
7. When you see
Start the
pump?
, press
´
.
NOTE:
• If you put the battery in backwards, the display will remain
m Match
attery»
Stop the Pump?
Press Y or N
pen
on»
Press
Battery
Door
Button
Start the Pump?
Press Y or N