Reservoir Volume
is zero
NEXT to continue
Reservoir Volume
Low
NEXT to continue
Upstream Occlusion
Press STOP to silence
(Model 6101 only)
Alarms and Messages
When you see:
Take this action:
You hear
9 beeps
: the infusion
profile has ended. Disconnect or
change the CADD
™
Administra-
tion set as appropriate.
The Reservoir Volume has
reached 0.0 ml. Press
„
to
stop the alarm. Then change the
fluid container if appropriate.
The Reservoir Volume value is
low, indicating that the level of
fluid in the fluid container is
low.
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed clamp,
or an air bubble in the tubing
between the fluid container and
pump. Press
⁄
to stop the
pump and silence the alarm for
2 minutes, then remove the
obstruction and press
„
to
restart the pump.