27
High Pressure
Pump Stopped
Air in line detected
Pump will not run
NEXT to silence
Delivery Stopped
(Model 6101 only)
Alarms and Messages
When you see:
Take this action:
There is air in the tubing or the
tubing is not threaded through
the air detector. Press
„
to
stop the alarm. Then follow your
clinician’s instructions for priming.
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed clamp,
or an air bubble in the tubing
between the fluid container and
pump. Press
⁄
to stop the
pump and silence the alarm for
2 minutes, then remove the
obstruction and press
„
to
restart the pump.
There may be a kink in the
tubing or a clamp may be
closed. Unkink the tubing or
open the clamp and the pump
will resume delivery. You may
press
⁄
to stop the pump
and silence the alarm for 2
minutes. After you remove the
cause of high pressure, start the
pump if necessary. If the alarm
continues, contact your clinician.