17
Stopping and Restarting the Pump
Stopping the pump stops delivery of medication. Whenever
the pump is stopped, the amber light will blink.
• If you stop the pump
during the Infusion Profile
, delivery
will stop, but will continue from the same point when you
restart the pump.
• If you stop the pump
during KVO delivery
, the Infusion
Profile and the Reservoir Volume will automatically be
reset. This means that when you restart the pump, it will
start from the beginning of the Infusion Profile.
1. Press
⁄
.
2. When
Stop the Pump?
appears, press
´
.
You will see two screens if you
stopped the pump during KVO
delivery:
STOPPED
will appear on the
main screen, and the amber
light will blink.
Stop the Pump?
Press Y or N
*** TPN 6230X ***
STOPPED
Press NEXT to advance
Infusion Profile
has been reset
Reservoir Volume
3000.0 ml
<Resetting...>