WD1050
12/04/2010
Pag. 29
Alarm ID
Description
Action
32
Water in chamber - backwater.
Make a Reset cycle. Repeat the cycle.
Call the Technical Assistance Service
if the fault persists.
33
Lack of water in the steam condenser.
Indicates that there is no water in the
steam condenser when there should
be due to sprayer nozzle activation.
Make sure the solenoid valve input to
the condenser activates and operates
correctly. Also make sure that the
drain pump does not remain activated
permanently.
34
Condenser drainage failed.
Intervenes if "SLC" (work level sensor
of the condenser) does not deactivate
after 120 seconds from drain pump
activation.
Make sure the drain pump operates
correctly. Also make sure that the
condenser drain respects the
recommended values and it not
obstructed.
Replace the level sensor, if necessary.
35
Condenser level switch failed
This alarm intervenes when the level
switch does not work: check the
electrical connections.
Call the Technical Assistance Service
36
Condenser draining pump failed.
Check if the condenser draining hose
has been correctly installed.
Call the Technical Assistance Service
52
Door electrically open.
Make sure that the door is properly
closed before beginning a cycle.
Check the conditions of the micro-
switches related to the closure. Call
the Technical Assistance Service if
the fault persists.
54
Door mechanically open.
Make sure that the door is properly
closed before beginning a cycle.
Check the conditions of the micro-
switches related to the closure. Call
the Technical Assistance Service if
the fault persists.
56
Doorlock fault.
Make sure that the door is properly
closed before beginning a cycle. Call
the Technical Assistance Service if
the fault persists.
73
Cycle Archive fault.
Error inside the microprocessor.
Call the Technical Assistance Service
if the fault persists.
77
Initial temperature inside the chamber greater
than 45°C; prewash phase must be carried out at
a temperature lower than 45°C.
The alarm appears (if the relative
option has been selected) if the
temperature of the water entering in
the first phase exceeds 45°C.
78
Microprocessor fault; this may happen under
some circumstances (i.e. voltage drops).
Error inside the microprocessor.
Call the Technical Assistance Service
if the fault persists.