WD1050
12/04/2010
Pag. 22
NOTE
If demi water is not available don't connect the demi water pipe to cold
or hot water, leave it disconnected.
The demi water circuit inside the machine don't pass through the built in
softener, so the demi water connection must be used only for
demineralised water.
When the machine is in stand-by, i.e. when no cycle is running, to set the demi water
parameter proceed as follows:
•
Press the ON/OFF button (3) to switch on the machine;
•
Turn the programme selector knob (1) to position 7;
•
Press the Door Open button (6) for 4 seconds to enter the demi water setup
mode;
•
On the display, it may appear "OE" when the demi water has been already set
or ‘–’ when it has not been set yet;
•
To set the demi water (demineralised water available), press "°C" (5) ;
•
To delete this setting (demineralised water not available), press "
►
ll" (4);
•
To terminate, turn the programme selector knob to position 7 and press the
Door Open button (6) for 4 seconds.
6.12. PRINTER
On the WD1050 a serial port (RS232) is available as a standard feature (see picture
below) and to which the SMEG printer, named “WD-PRINTE” with code 901783, may
be connected. The SMEG printer is capable of printing out the following data:
•
Serial Number of the WD1050 "Serial Number"
•
Number of cycles carried out "Cycle N."
•
Date and Time
•
Program Number indicating Thermal Disinfection Temperature and holding time
(for example 93°C 10’)
•
Detergent control
o
"Inflow Detergents P1 Deter" - for the detergent;
o
"Inflow Detergents P2 Neutr" - for the neutralizing agent;
•
Temperature control
o
‘TL1’: indicates the temperature measured inside the washing chamber.
It is the main temperature of the cycle;
o
‘TLC’: indicates the control temperature as requested by the EN15883
standard;
o
‘Target T’: stands for the target temperature value of the phase
•
Heating element
o
‘Res ON’: indicates that the heating element is on;
o
‘Res OFF’: indicates that the heating element is off;
•
Starting Time of the Cycle