LD303 - Operation and Maintenance Instruction Manual
3.28
The number of points is configured in parameter FACTORY_CURVE_LENGTH, being in the
maximum 5 points. The entry points will be configured in the FACTORY_CURVE_X and of output in
the FACTORY_CURVE_Y.
The Parameter FACTORY_CURVE_BYPASS controls the enabling/disabling of the curve and has
the following options:
x
"Disable ",
x
"Enable and Backup Cal " ,
x
"Disable and Restore Cal " ,
x
"Disable or Allows to enter the points"
To configure the points of the curve, the option "Disable or Allows to enter the points” must be
chosen. Then press the "Characterization Cal". The following message appears: "This Function
alters XMTR characteristics. Proceed? Y/N". To proceed, select "Yes". A new message appears: "Is
XMTR connected to accurate pressure standard ?". To proceed, select "Yes". Apply the desired
pressure and wait that the same one stabilizes. If the pressure is not stable, select "No-read again".
If it is stable, enter "Yes" and then, type the applied pressure P1. Repeat this procedure for the next
point P2. After that, if the user wants to configure more points, just repeat this procedure up to 5
points. If not, just select "No" for the question “Do you want to configure more points? ".
After configuring the points, the user needs to qualify the curve. The option "Enable and backup
cal", enables the curve and save the calibration settings. The option "Disable and restore the cal",
disables the curve and restores the calibration settings. The option “Disable”, just disables the curve
and does not take care about the calibration settings.
Figure 3.29 – The Characterization Curve Configuration
The Characterization Curve can have a minimum of 2 and up to 5 points. These points should be
between the calibrated range for better results.
By the list box the
user can enable or
disable the
Characterization
Curve, enter the
points, restore or
backup the curve
entered.
After choosing the
condition for the curve,
please, press this key to
initiate the characterization
curve method.
Select this page to see the
configured points.
Содержание LD303
Страница 1: ...Profibus PA Pressure Transmitter L D 3 0 3 M E...
Страница 8: ...LD303 Operation and Maintenance Instruction Manual VIII...
Страница 15: ...Installation 1 7 Figure 1 1 f Dimensional Drawing and Mounting Position Sanitary Transmitter with Extension...
Страница 22: ...LD303 Operation and Maintenance Instruction Manual 1 14...
Страница 26: ...LD303 Operation and Maintenance Instruction Manual 2 4...
Страница 49: ...Configuration 3 23 Figure 3 24 Block Mode for Totalizer Block Figure 3 25 Set Reset for Totalizer Block...
Страница 74: ...LD303 Operation and Maintenance Instruction Manual 4 6 Figure 4 4 Exploded View...
Страница 83: ...Maintenance Procedures 4 15 Figure 4 8 TPE Software Screen...
Страница 102: ...LD303 Certifications Information A 4 CSA NEMKO and EXAM CEPEL NEPSI...
Страница 104: ...LD303 Certifications Information A 6 CEPEL...
Страница 105: ...Appendix A A 7 Control Drawing Factory Mutual FM...
Страница 106: ...LD303 Certifications Information A 8...
Страница 108: ...LD303 Operation and Maintenance Instruction Manual B 2...