5.4
Technical Characteristics of High Performance - CODE L1
High Performance option (code L1) is available under the following conditions only:
Application
Differential and Gage
Range
D2
D3
D4
M2
M3
M4
-50
-250
-2500
-50
-100
-100
to
to
to
to
to
to
50 kPa
250 kPa
2500 kPa
50 kPa
250 kPa
2500 kPa
-200
-36
-360
-200
-14.5
-14.5
to
to
to
to
to
to
200 inH
2
O
36 psi
360 psi
200 inH
2
O
36 psi
360 psi
Diaphragm
Material
316L SST
Hastelloy C276
Fill fluid
Silicone
Performance Specifications
Reference
Conditions
Span starting at zero, temperature of 25 °C (77 °F), atmospheric pressure, power supply of 24 Vdc, silicone oil fill fluid,
isolating diaphragms in 316L SST and digital trim equal to lower and upper range values.
Accuracy
RANGE 2:
0.2 URL
span
URL:
± 0.04% of span
0.05 URL
span < 0.2 URL:
± [0. 0.003667 URL/span]% of span
0.0085 URL
span < 0.05 URL:
± [0.0021 + 0.004645 URL /span]% of span
RANGES 3 and 4:
0.1 URL
span
URL:
± 0.05% of span;
0.05 URL
span < 0.1 URL:
± [0.005 + 0.045 URL/span]% of span
0.0085 URL
span < 0.05 URL:
± [0.0021+ 0.004645 URL/span]% of span
Stability
For range 2:
± 0.05% of URL for 6 months
For range 3:
± 0.075% of URL for 12 months
For range 4:
± 0.1% of URL for 24 months
± 0.2% of URL for 12 years, at 20 °C temperature change and up to 7 MPa (1000 psi) {70 bar} of static pressure,
environment free of hydrogen migration.
Temperature
Effect
From -10 °C to 50 °C
,
protected from direct sun radiation:
0.2 URL
span
URL:
± [ 0.018% URL + 0.012% span] per 20 °C (68 °F)
0.0085 URL
span < 0.2% URL:
± [0.002% URL + 0.002% span] per 20 °C (68 °F)
Static Pressure
Effect
Zero error:
± 0.025% URL per 7MPa (1000 psi)
The zero error is systematic and can be eliminated by calibrating at the operating static pressure.
Span error:
Correctable to ± 0.2% of reading per 7 MPa (1000 psi)
NOTES
Hasteloy is a trademark of the Cabot Corp.
Monel is a trademark of International Nckel Co.
Viton and Teflon are trademarks of E. I.
DuPunt de Nemours & Co
Fluorolube is a trademark of Hooker Chemical Corp.
Halocarbon is a trademark of Halocarbon.
Profibus is a trademark of Profibus International
Smar Pressure Transmitters are protected by US
patent number 6,433,791
Содержание LD303
Страница 1: ...Profibus PA Pressure Transmitter L D 3 0 3 M E...
Страница 8: ...LD303 Operation and Maintenance Instruction Manual VIII...
Страница 15: ...Installation 1 7 Figure 1 1 f Dimensional Drawing and Mounting Position Sanitary Transmitter with Extension...
Страница 22: ...LD303 Operation and Maintenance Instruction Manual 1 14...
Страница 26: ...LD303 Operation and Maintenance Instruction Manual 2 4...
Страница 49: ...Configuration 3 23 Figure 3 24 Block Mode for Totalizer Block Figure 3 25 Set Reset for Totalizer Block...
Страница 74: ...LD303 Operation and Maintenance Instruction Manual 4 6 Figure 4 4 Exploded View...
Страница 83: ...Maintenance Procedures 4 15 Figure 4 8 TPE Software Screen...
Страница 102: ...LD303 Certifications Information A 4 CSA NEMKO and EXAM CEPEL NEPSI...
Страница 104: ...LD303 Certifications Information A 6 CEPEL...
Страница 105: ...Appendix A A 7 Control Drawing Factory Mutual FM...
Страница 106: ...LD303 Certifications Information A 8...
Страница 108: ...LD303 Operation and Maintenance Instruction Manual B 2...