Sirona Dental Systems GmbH
Operating Instructions CEREC AC
8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
63 33 582 D3492
D3492.201.01.22.02 10.2018
57
6.
Choose a container. Find an example on the left.
7.
Cautiously fill the container to a filling level of 20mm - 25mm
(13/16in - 1in) under the small hole (A) of the prismatic tube when
immersed. A funnel can be used to assist with filling the liquid. Do
not spill any disinfectant. If spillage occurs please follow disinfectant
´s manufacturer´s safety indications for cleaning.
NOTE
Only place the prismatic tube into the container and not the entire
CEREC Bluecam.
Ensure that the prismatic tube is placed upright in the disinfectant
(glass down) so that the solution cannot penetrate the inside of the
prismatic tube.
8.
Insert the prismatic tube in the correct position (glass down).
9.
Leave the prismatic tube at least for 12 minutes for CIDEX
®
OPA or
30 minutes for Sporox II in the disinfectant. While exceeding the
recommended immersion time may not cause damage to the
prismatic tube, the prismatic tube should be removed as soon as
possible after the recommended time.
10. Remove the prismatic tube from the container.
11. Thoroughly rinse the prismatic tube with tap water of potable water
quality that meets Federal Clean Water Standards for at least 30
seconds. Ensure that no water is able to penetrate into the interior
of the mirror sleeve.
12. Wipe off the prismatic tube. Use a cotton gauze moistened with tap
water of potable water quality that meets Federal Clean Water
Standards for this purpose.
13. Dry the prismatic tube using a clean soft, lint-free cloth.
14. Store the prismatic tube in such a way that it is protected from
contamination until the next use.
15. Remove the protective cap from the camera before use.
16. Carefully re-attach the prismatic tube (see "Refitting the prismatic
tube [→ 53]“).