6 Installation and startup
6.4 Initial startup
Sirona Dental Systems GmbH
Operating Instructions CEREC AC
34
6.4.7
Changing from right-handed to left-handed operation
In the factory default setting, the left button trackball button corresponds
to a foot control entry. If you would like to change this assignment to the
right trackball button, your CEREC service technician can do this for
you.
6.4.8
Switching the units on
NOTE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
ü
Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry (for at least one hour)
Ä
The unit is dry and can be put into operation.
CAUTION
Use only the supplied power cord
Use only the power cord supplied by Sirona to connect the acquisition
unit to the power supply.