Sirona Dental Systems GmbH
Operating Instructions CEREC AC
6 Installation and startup
6.4 Initial startup
63 33 582 D3492
D3492.201.01.22.02 10.2018
31
6.4.3
Line voltage
230
115
D
A
B
C
A
Voltage selection insert
C
Fuse module
B
Main fuses
C
Window
➢ Check the set line voltage. The value of the line voltage must be
visible in the window with the module inserted (230V in Europe,
100V in Japan and 115V in the USA). If the set voltage does not
agree with the actual line voltage, you must change this setting:
WARNING
Risk of electric shock
Electric shock due to inserted power plug.
➢ Disconnect the power plug before selecting the line voltage.
ü
The line voltage can be switched from 230V to 100V or 115V and
vice versa.
230
115
230
115
100
1.
To do this, unlatch the fuse module with a screwdriver and pull the
module out.