8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
63 33 582 D3492
D3492.201.01.22.02 10.2018
61
8.1.5.2.6
Pre-cleaning of the prismatic tube
Cleaning process immediately after using the CEREC Bluecam: While
the prismatic tube is attached to the camera, wipe off so that any
surface contamination cannot harden and adhere to the surface of the
prismatic tube.
Use a cotton gauze moistened with tap water of potable water quality
that meets Federal Clean Water Standards and pH neutral soap or use
a clean lint-free cloth which has first been soaked in one of the cleaning
products listed in the „Cleaning and disinfection agents [→ 50]“ section
to remove visible contamination.
Then dry the prismatic tube with a clean lint-free cloth.
8.1.5.2.7
Disinfecting the camera body
NOTE
Do not spray the CEREC Bluecam or immerse it in cleaning agents or
disinfectants!
➢ Use a new cotton gauze moistened with one of the cleaning
products listed in the „Cleaning and disinfection agents [→ 50]“
section. Wipe the camera body first and then the prismatic tube.
This step disinfects the camera body and removes cleaning
residues from the prismatic tube. Then dry the prismatic tube with a
clean lint-free cloth.
8.1.5.2.8
High-level disinfection of the prismatic tube with sapphire glass
The complete process for high-level disinfection is as follows:
With HLD set
A HLD set to support the HLD process should be ordered from Sirona
with REF 63 46 907.
1.
Remove the CEREC camera support (if present).
2.
Pre-cleaning (see „Pre-cleaning of the prismatic tube [→ 61]“).
3.
Remove the prismatic tube from the CEREC Bluecam (see
“Removing the prismatic tube [→ 59]”).
4.
Place the protective cap on the camera and place the camera in the
camera cradle.
5.
Use one of the following disinfectants for the high-level disinfection:
CIDEX
®
OPA, Sporox II.
CAUTION
Observe the disinfectant‘s manufacturer‘s safety indications!