User manual - DATA RECORDER SRD-99 and cooperating software
This is a class A unit. In a residential or a similar area it can cause radio
frequency interference. In such cases the user can be requested to use
appropriate preventive measures.
Carefully check that the isolation used with the unit (
) meets the
expectations and if necessary use appropriate measures of overvoltage protection.
Additionally, insure the appropriate air and surface insulation gaps when installing
(
Range of operating voltage
Air gap [mm]
Surface gap [mm]
Up to 50V RMS or DC
0,2
1,2
Up to 100V RMS or DC
0,2
1,4
Up to 150V RMS or DC
0,5
1,6
Up to 300V RMS or DC
1,5
3,0
Tab. 3.1. Minimal insulation gaps according to EN 61010-1
Fig. 3.1. Schematic diagram showing the insulation between individual circuits of the unit.
9
!
!
Internal circuits
relay 2
relay 1
External sensor
supply output
Measurement inputs
RS 485
interface
Reinforced insulation 1min @ 2300V AC
Supplementary insulation 1min @ 1350V AC
Basic insulation 1min @ 500V AC
No insulation
Relay circuits
Power
supply