Technical data
TP 28 1300 E
- 54 -
© SIKA
•
Ba_TP281300E
•
08/2015
9.2
Heating up and cooling times
The following measurement conditions apply to heating and cooling times:
•
All times are elapsed times without settling time.
•
The measurements were carried out at a room temperature of approx. 23 °C
•
Without cap, without cover
•
Controlled by internal reference sensor.
Dry block series TP 28 1300 E:
•
Reference sensor thermocouple TYPE S (
∅
6 mm), ceramic immersion tube.
•
The reference sensor sits in the middle of the adapter sleeve at full depth.
Heating times:
Temperature change
Duration
400 °C to 1300 °C
90 min
400 °C to 600 °C
90 min
600 °C to 800 °C
80 min
800 °C to 1000 °C
75 min
1000 °C to 1200 °C
70 min
Cooling times:
Temperature change
Duration
1300 °C to 1000 °C
20 min
1000 °C to 800 °C
20 min
800 °C to 600 °C
30 min
600 °C to 400 °C
50 min