TP 28 1300 E
Device description
Technical changes reserved
- 33 -
Scope of delivery and accessories:
Scope of delivery
Accessories (optional)
TP 28 1300 E
according to order data.
Test certificate.
Mains cable.
Sleeve exchange tool.
Operating manual.
Protective packaging / transport protec-
tion.
Transport case.
Adapter sleeve(s).
DAkkS certificate.
Works certificate.
Converter USB.
PC software.
IMPORTANT!
Use the type plate to check if the delivered unit corresponds to your order.
In particular, for devices with electrical components, check to see if the correct pow-
er supply voltage is specified.
1.2
Intended use
The calibrators of the TP 28 1300 E may only be used for testing of temperature switches,
thermostats, thermocouples, thermometers and resistance thermometers.
The calibrators may not be used for warming up or heating other parts or gases.
Ensure that the test temperature is suitable for the test specimen to prevent the test speci-
men from being destroyed.
The calibrators have been designed for indoor use only.
The operational safety of the device supplied is only guaranteed by intended use. The speci-
fied limits (
§ 9 "Technical data") may under no circumstances be exceeded.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
Never touch the metal block, the adapter sleeve or the test specimen at temperatures
above 35 °C or below 10 °C.
Allow the calibrator to cool before you remove the test specimen, change the adapter
sleeve or the calibration insert or switch off the device.