Signametrics
16
Accuracy ± (% of r Amps)
Range
Frequency [1]
24 hours
23
C
1
C
90 Days
23
C
10
C
One Year
23
C
10
C
10 Hz - 20 Hz
3.8 + 4
A
2.7 + 4
A
2.9 + 4
A
20 Hz - 47 Hz
0.9 + 4
A
0.9 + 4
A
1.0 + 4
A
47 Hz - 1 kHz
0.04 + 1.5
A
0.08 + 3
A
0.12 + 4
A
2.4 mA
1 kHz - 10 kHz
0.12 + 4
A
0.14 + 4
A
0.22 + 4
A
10 Hz - 20 Hz
1.8 + 30
A
2.6 + 30
A
2.8 + 30
A
20 Hz - 47 Hz
0.6 + 30
A
0.9 + 30
A
1.0 + 30
A
47 Hz - 1 kHz
0.07 + 10
A
0.15 + 20
A
0.16 + 30
A
24 mA
1 kHz - 10 kHz
0.21 + 30
A
0.3 + 40
A
0.4 + 40
A
10 Hz - 20 Hz
1.8 + 400
A
2.7 + 400
A
2.8 + 400
A
20 Hz - 47 Hz
0.6 + 400
A
0.9 + 400
A
1.0 + 400
A
47 Hz - 1 kHz
0.1 + 100
A
0.17 + 180
A
0.2 + 220
A
240 mA
1 kHz - 10 kHz
0.3 + 300
A
0.35 + 350
A
0.4 + 400
A
10 Hz - 20 Hz
1.8 + 4 mA
2.5 + 4.5 mA
2.7 + 5 mA
20 Hz - 47 Hz
0.66 + 4 mA
0.8 + 6 mA
0.9 + 6 mA
47 Hz - 1 kHz
0.3 + 3.8mA
0.33 + 3.8 mA
0.35 + 4 mA
2.4 A
1 kHz - 10 kHz
0.4 + 4mA
0.45 + 4.5 mA
0.5 + 5 mA
[1] All AC Current ranges have typical measurement capability of at least 20 kHz.
2.6 Leakage Measurement (SMU2064)
Characteristics
Burden Voltage:
< 100
V
Test Voltage:
Adjustable between -10V to +10V in 5mV steps
Accuracy ± (% of r Amps) [1]
Range Full
Scale
6-½ Digits
Resolution
24 hours
23
C
5
C
90 Days
23
C
5
C
One Year 23
C
5
C
240
A 240.0000
A
0.1 pA
0.07 + 40pA
0.1 + 45pA
0.17 + 60pA
2.4
A 2.400000
A
1 pA
0.05 + 70pA
0.08 + 90pA
0.21 + 150pA
24
A 24.00000
A
10 pA
0.05 + 400pA
0.08 + 600pA
0.13 + 0.8nA
[1] With Aperture set to
≥
0.5 Sec, and within one hour from Zero (Relative control).
2.7 RTD Temperature Measurement
Ro:
Variable 10
to 10 k
Measurement Method:
4-Wire
Temperature units:
Selectable
o
C or
o
F
RTD Type
Ro (
)
Resolution Temperature
range
Temperature Accuracy 23
C
5
C [1]
One Year
pt385, pt3911,
pt3916, pt3926
100, 200
0.01
C
-150 to 650
C
0.06
C
pt385, pt3911,
pt3916, pt3926
500, 1 k
0.01
C
-150 to 650
C
0.03
C
Cu (Copper)
Less than 12
0.01
C
-100 to 200
C
0.18
C for temperatures
20
C,
0.05
C
otherwise
Cu (Copper)
Higher than 90
0.01
C
-100 to 200
C
0.10
C for temperatures
20
C,
0.05
C
otherwise
[1] With Aperture of 0.5s and higher, using a 4-wire RTD. Measurement accuracy does not include RTD probe
error.
Содержание SMU2060
Страница 26: ...Signametrics 26 PXI Instrumentation Switching modules SMX4030 SMX4032 IVI COM driver...
Страница 81: ...81 Signametrics Postive Value 100 Warning code Example char cBuf 64 int status status DMMGetCalDate 0 cBuf...