135
Signametrics
Remarks
This function sets the Aperture to an integer multiple,
iMultiple
, of the specified power
line cycle. The line frequency, iLineFreq, can be 50Hz, 60Hz or 400Hz. The multiple
range can be 1 to 50. Also see DMMSetAperture().
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
iLineFreq
int
Identifies the powr line frequnecy. Can take a value of 50, 60 or
400.
bMultiple
int
Defines the Aperture value as a multiple of power line cycles.
Can be set between 1 and 50.
Return Value
The return value is one of the following constants.
Value Meaning
DMM_OKAY
Function succeeded.
Negative Value
Error code
Example
status = DMMSetPLC(0, 60, 10); /* Set aperture to 166.667ms */
DMMSetPulseGen
2060
2064
Description
Set the parameter of the pulse generator output.
#include "SMX2060.h"
#include "DMMUser.H"
int
DMMSetPulseGen
(
int
nDmm,
double
dPwidth,
double
dNwidth,
int
nPulses
)
Remarks
This function sets the parameters of the pulse generator source.
dPwidth
sets the positive,
or active width portion of the of the pulse, d
Nwidth
sets the negative (0V) portion of the
pulse.
nPulses
sets the number of pulses to be generated, as well as the mode. Both
d
Pwidth
and
dNwidth
are in seconds.
dPwidth
and
dNwidth
can be set between 25µ
(25.0e-6) and 3s (3.0). The value of
nPulses
can be set between 0 to 32,000. A value of
0 sets the pulse generator to a free running mode. Other values set the number of pulses
in a burst. The DMM must be set to the
PULSE_GEN
function prior to using this
function. The
DMMSetDCVSource
function controls the amplitude of the pulse (-10V
to +10V) while in pulse generator mode. If either dNwidth or dPwidth are greater than
0.0655s, the width resolution becomes 100us, and the minimum value of dNwidth and/or
dPwidth is 1.5ms. This function requires Driver version 1.60 and Microcode version 1.29
or higher.