Signametrics
Example
int i =
WrCalFileToStore
(0,
"C:\\SM60CAL.dat"
)
;
WrCalStoreToFile
SMU2060
SMU2064
Description
Transfer the contents of the on-board cal store to a file.
#include "SMU2060.h"
int WrCalStoreToFile (int
nDmm,
LPCSTR
lpszCal,
int
mode
)
Remarks
This function copies the calibration record stored in the on-board none volatile memory
of the DMM to the specified calibration file, pointed to by lpszCal. If
mode
is ‘a’ and a
file exists, the record is appended to the end of this file. If
mode
is ‘w’, a new file is
created, wiping out the old if it exists.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
mode
int
Sets the file creation mode.
lpszCal
LPCSTR
Points to the name of the file containing the calibration
constants for the DMM. Calibration information is normally read
from the file named
SM60CAL.DAT
located in the C:\ root directory.
Return Value
Integer error code.
Value Meaning
Any value not 0 int
Eror or warning code
Example
int i =
WrCalStoreToFile
(0,
"C:\\SM60CAL.dat", ‘a‘
)
;
DMMGetSupplyV
SMU2060
SMU2064
Description
Returns the one of the DMM supplies voltages.
#include "SMU2060.H"
int
DMMGetSupplyV
(
int
nDmm,
, double *
lpdVoltage
)
Remarks
This function makes a measurement of one of the DMM power supplies voltages as an
indication of the USB supply voltage level. The nominal value is -12V. The USB
interconnect and some off the shelve hubs can make this voltage higher or lower than is
required. The acceptable value should be -10.5 to -13.5V. Voltages higher than -13.5V
may damage the SMU2055 and voltages below -9.0 are inadequate for proper operation,
and is usually indicative of poor USB cable. The value of this voltage is stored at a
double precision location pointed to by
lpdVoltage
.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
lpdVp;tage
LPSTD
Points to a double to hold the result.