Signametrics
18
Range Input
Typical
Measurement
Time [1]
Typical Measurement
speed (rps) [1]
1,200 pF
5% of Scale
19.5 ms
51.3
1,200 pF
Full Scale
52.3 ms
19.1
12
F
5% of Scale
70.0 ms
14.3
12
F
Full Scale
118ms 8.5
120
F
5% of Scale
8.9 ms
112.4
120
F
Full Scale
127 ms
7.9
1.2
F
5% of Scale
15.6 ms
64.1
1.2
F
Full Scale
175 ms
5.7
12
F
5% of Scale
14.1 ms
70.9
12
F
Full Scale
480 ms
2.1
120
F
5% of Scale
17.3 ms
57.8
120
F
Full Scale
50.3 ms
19.9
1.2 mF
5% of Scale
52.6 ms
19.0
1.2 mF
Full Scale
151.5 ms
6.6
12 mF
5% of Scale
52.8 ms
18.9
12 mF
Full Scale
170 ms
5.9
[1] This time depends on the value measured capacitance. The SMU2060 is about 10 times slower than the
SMU2064.
2.9.3 Capacitance, In-Circuit Method (SMU2064)
Method
Variable frequency AC
Adjustable Peak Voltages Stimulus
100mV to 5.0V
Parallel Load Resistance
as low as 100
Accuracy ± (% of r Farads) [1]
Range Full
Scale
3-½
Digits
Resolution
One Year
23
C
5
C [2]
24
F 23.99
F
10 pF
5 ± 200 pF
240
F 239.9
F
100 pF
5 ± 1
F
2.4
F 2.399
F
1000 pF
3 ± 5
F
24
F 23.99
F 10
F
3 ± 50
F
240
F 239.9
F 100
F
5 ± 500
F
2.4 mF
2.399 mF
1
F
6 ± 5
F
[1] Within one hour of AC Caps Open Cal operation, and relative correction.
[2] Specified for values higher than 5% of the selected range with Aperture > 0.2s
2.9.4 Inductance Measurement (SMU2064)
Accuracy ± (% of r inductance) [1]
Range
Test frequency
Full Scale
4 ½ Digits
Resolution
Accuracy 23
C
5
C
One Year [2]
33
H
100 kHz
33.000
H
1
H
3.0% + 500
H
330
H
50 kHz
330.00
H 10
H
2.0% + 3
H
3.3 mH
4 kHz
3.3000 mH
100
H
1.5% + 25
H
33 mH
1.5 kHz
33.000 mH
1
H
1.5% + 200
H
330 mH
1 kHz
330.00 mH
10
H
2.5 + 3 mH
3.3 H
100 Hz
3.3000 H
100
H
3 + 35 mH
[1] Within one hour of Zero, and Open Terminal Calibration.
[2] Accuracy is specified for values greater than 5% of the selected range.