![Signametrics Corporation SMU2060 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/signametrics-corporation/smu2060/smu2060_operators-manual_1268488062.webp)
Figure 4.26. A load cell application maintains common mode limits on I+ and I-.
The table below (Fig. 4-28) lists which of the above auxiliary DCV terminals and measurement functions
are available for use during various measurement and sourcing operations. It is important to adhere to the
following since an unavailable terminal implies that is likely to be shorted to the V,
Ω
- terminal.
SMX2064 Function DMMReadHiSense()
DMMReadLoSense() DMMReadHiLoSense()
VDC
Diode Test
2-Wire Ohms
VDC Source
VAC Source
IDC Source
Leakage
Extended Ohms
Caps
Inductance
Figure 4.28. Auxiliary DCV measurement functions availability during various DMM operations.
Signametrics
62