DMMSetTrimDAC
SMU2060
SMU2064
Description
Set the Trim DAC level.
#include "SMU2060.h"
#include "USBDMMUser.h"
int
DMMSetTrimDAC(int
nDmm
, int
iValue
)
Remarks
This function sets the Trim DAC to a value between 0 and 100. The trim DAC can be set
to augment the main 12 bit DAC, whenever it is not automatically performed, such as in
VDC and VAC source while
OPEN_LOOP
mode is selected. An example would be in
DCI source, or when setting the Comparator Threshold. This function consumes a lot of
the on-board microcontroller’s resources and must be turned off when not in use. Use
DMMDisableTrimDAC
to turn it off. With the Trim DAC the effective resolution of the
composite DAC is increased to 16 bits. See
DMMSetDCVSource
and
DMMSetACVSource
for more details.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
iValue
int
Amplitude can be set from 0 to 100, corresponding to 0% to
100% Trim DAC level.
Return Value
Integer error code.
Value Meaning
DMM_OKAY
Operation successfully completed.
Negative Value
Error code
Example
DMMSetDCVSource(0, 5.0); // Set source to 5V
DMMSetTrimDAC(0, 50); // add about 2.5mV to output
DMMStartTotalizer
SMU2060
SMU2064
Description
Clear the totalized value and start the totalizer.
#include "USBDMMUser.h"
#include "SMU2060.h"
int
DMMStartTotalizer
(
int
nDmm,
int
Edge
)
Remarks
To use this function the DMM must be in ACV measurement function, and a valid range
must be selected. This function clears the Totalized count, sets the edge sense, and starts
the Totalizer. The totalized value can be read during the accumulation period. However,
it could affect the count by the interruption. If no reads are performed during
accumulation, the input rate can be as high as 30,000 events per second. If reads are
performed during the accumulation period, this rate could be as low as 20,000 events per
second. The Threshold DAC sets the threshold at which signals are counted. During
accumulation, no other command (except
DMMReadTotalizer
) should be used. When
done, this function must be turned off using
DMMStopTotalizer
. After the Totalizer is
stopped, the accumulated result can be read using
DMMReadTotalizer
. A normal
procedure would be to set the DMM to the ACV function, select voltage range, set the