Signametrics
14
Accuracy ± (% of r Volts) [1]
Range
Frequency
24 hours
23
C
1
C
90 Days
23
C
5
C
One Year
23
C
5
C
350 Hz - 800 Hz
0.6 + 150
V
0.65 + 170
V
0.7 + 200
V
800 Hz - 10 kHz
0.13 + 100
V
0.14 + 110
V
0.15 + 120
V
10 kHz - 50 kHz
0.55 + 160
V
0.6 + 200
V
0.63 + 230
V
240 mV
50 kHz - 100 kHz
5.3 + 350
V
5.4 + 370
V
5.6 + 400
V
350 Hz - 800 Hz
0.93 + 1.3 mV
0.96 + 1.5 mV
1.0 + 1.7 mV
800 Hz - 10 kHz
0.068 + 1 mV
0.075 + 1.1 mV
0.08 + 1.2 mV
10 kHz - 50 kHz
0.62 + 1.2 mV
0.65 + 1.3 mV
0.70 + 1.5 mV
2.4 V
50 kHz - 100 kHz
5.1 + 1.5 mV
5.2 + 1.7 mV
5.3 + 2 mV
350 Hz - 800 Hz
0.93 + 12 mV
0.96 + 14 mV
1.0 + 16 mV
800 Hz - 10 kHz
0.065 + 10 mV
0.068 + 11 mV
0.073 + 13 mV
10 kHz - 50 kHz
0.31 + 18 mV
0.33 + 21 mV
0.35 + 25 mV
24 V
50 kHz - 100 kHz
2.0 + 30 mV
2.2 + 35 mV
2.4 + 40 mV
350 Hz - 800 Hz
0.93 + 120 mV
0.96 + 130 mV
1.0 + 150 mV
800 Hz - 10 kHz
0.062 + 100 mV
0.065 + 110 mV
0.08 + 130 mV
10 kHz - 50 kHz
0.32 + 150 mV
0.4 + 170 mV
0.45 + 200 mV
240 V
50 kHz - 100 kHz
2.5 + 200 mV
2.8 + 240 mV
3.2 + 300 mV
350 Hz - 800 Hz
1.0 + 180 mV
1.1 + 200 mV
1.1 + 250 mV
800 Hz - 10 kHz
0.065 + 150 mV
0.07 + 200 mV
0.08 + 230 mV
10 kHz - 50 kHz
0.34 + 200 mV
0.45 + 250 mV
0.5 + 300 mV
330 V
50 kHz - 100 kHz
2.5 + 270 mV
2.8 + 350 mV
3.2 + 400 mV
[1] With Aperture set to
≥
0.16 Sec
2.4.2 AC Peak-to-Peak Measurement (SMU2064)
Measures the peak-to-peak value of a repetitive waveform.
ACV
Range
Lowest specified
input voltage (Vp-p)
Full Scale [2]
reading (Vp-p)
Resolution
Typical Accuracy 23
C
5
C
One Year [1]
240 mV
0.1 V
1.900 V
1 mV
0.5
3 mV
2.4 V
1.0 V
16.00 V
10 mV
0.5 ± 40 mV
24 V
10 V
190.0 V
100 mV
0.5 ± 700 mV
240 V
100 V
850.0 V
1 V
0.55 ± 6 V
[1] Signal frequency range 30 Hz to 60 kHz.
[2] USB power level greatly effects full scale reading.
2.4.3 AC Crest Factor Measurement (SMU2064)
Measures the crest factor (CF) of a repetitive waveform
ACV
Range
Lowest specified
input voltage
(Vp-p)
Highest specified input
voltages (Vp-p)
Resolution
Typical Accuracy 23
C
5
C
One Year [1]
240 mV
0.1 V
1.9 V
0.01
2.2
0.3
2.4 V
1.0 V
16 V
0.01
2.1 ±0.1
24 V
10 V
190 V
0.01
2.0 ±0.1
240 V
100 V
700 V
0.01
2.0 ±0.1
[1] Crest factor measurement requires signal frequency of 30 Hz to 60 kHz.
Содержание SMU2060
Страница 26: ...Signametrics 26 PXI Instrumentation Switching modules SMX4030 SMX4032 IVI COM driver...
Страница 81: ...81 Signametrics Postive Value 100 Warning code Example char cBuf 64 int status status DMMGetCalDate 0 cBuf...
Страница 106: ...Signametrics 106 Postive Value 100 Warning code Example char cBuf 64 int status status DMMPeriodStr 0 cBuf...