SDM3055/SDM3055A 32
2.5.1 CALCulate:AVERage[:STATe]{ON|1|OFF|0}
CALCulate:AVERage[:STATe]?
Enables or disables statistics computation.
Parameter
Typical Return
{ON|1|OFF|0}, default OFF
0 (OFF) or 1 (ON)
Return the statistics of 100 frequency measurements:
CONF:FREQ
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
CALC:AVER:ALL?
Typical Response:
-4.10466677E-04,+3.13684184E-04,+1.75743178E-02,-6.74799085E-04
Statistics are cleared when the measurement function changes or when any of these
commands is executed:
CALCulate:AVERage:STATe ON
CALCulate:AVERage:CLEar
INITiate
MEASure:<
function
>?
READ?
*RST
SYSTem:PRESet
The instrument turns this setting OFF when the measurement function is changed or
after *RST or SYSTem:PRESet.