2.11.3 TRIGger:DELay:AUTO {ON|1|OFF|0} TRIGger:DELay:AUTO?
Disables or enables automatic trigger delay.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0(OFF)or1(ON)
Return 5 DC voltage measurements, with an automatic delay between each
measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL:AUTO 1
READ?
Typical Response
:
+3.07761360E-03,-1.16041169E-03,+5.60585356E-06,+1.21460160E-04,+2.85898
531E-04
Selecting a specific trigger delay using TRIGger:DELay disables the automatic trigger
delay.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).