Remote Manual
SIGLENT
SDM3055/SDM3055A 44
2.7 DATA Subsystem
This subsystem allows you to configure and remove data from reading memory. The
instrument clears all measurements from reading memory when the measurement
configuration changes, or when any of these commands is executed:
INITiate
MEASure:<
function
>?
READ?
*RST
SYSTem:PRESet
Command Summary
DATA:LAST?
DATA:POINts?
DATA:POINts:EVENt:THReshold
DATA:REMove?
2.7.1 DATA:LAST?
Returns the last measurement taken. You can execute this query at any time, even during
a series of measurements.
Parameter
Typical Return
(
none
)
One measurement with units. If no data is available, 9.91E37 (Not a
Number) is returned with units
Example
:
-4.79221344E-04 VDC
Return the last measurement
:
DATA:LAST?