2.8.4 MEASure:{FREQuency|PERiod}?
Sets all measurement parameters and trigger parameters to their default values for
frequency or period measurements and immediately triggers a measurement. The results
are sent directly to the instrument's output buffer.
Parameter
Typical Return
(
none
)
+7.1901
configure and read the default frequency measurements
:
MEAS:FREQ?
If the input voltage is too large for the selected voltage range (manual ranging), the
instrument displays the word Overload on the front panel and returns "9.9E37" from
the remote interface. Autoranging can be enabled for the input voltage.
2.8.5 MEASure:{RESistance|FRESistance}?
[{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement and trigger parameters to their default values for 4-wire resistance
(FRESistance) or 2-wire (RESistance) measurements, and immediately triggers a
measurement. The results are sent directly to the instrument's output buffer.Also specifies
the stalls through the incoming parameters.
Parameter
Typical Return
<range>
:
{200 Ω|2 kΩ|20 kΩ|200 kΩ|2 MΩ|10
MΩ|100 MΩ}
,
AUTO
(default) or DEFault
(
none
)
Configure 4-wire resistance measurements using the 200 Ω range with
default resolution. Then make and read one measurement:
:
MEAS:FRES? 200
Typical Response
:
+6.7101
Содержание SDM3055A
Страница 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Страница 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 113: ......
Страница 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...
Страница 140: ...Remote Manual SDM3055 SDM3055A Digital Mulimeter RC06035 E01A 2014 SIGLENT TECHNOLOGIES CO LTD ...