SDM3055/SDM3055A 40
2.6.6 CONFigure:{RESistance|FRESistance}
[{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for 4-wire
(FRESistance) or 2-wire (RESistance) resistance measurements.Also specifies the range
and resolution.
Parameter
Typical Return
<range>
:
{200 Ω|2 kΩ|20 kΩ|200 kΩ|2 MΩ|10
MΩ|100 MΩ}
,
AUTO(default)
(
none
)
Configure 4-wire resistance measurements using the 200 Ω range with default
resolution.Make and read two measurementsCONF:FRES 200
SAMP:COUN 2
READ?
Typical Response
:
+6.7101,+6.8301
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
Use READ? or INITiate to start the measurement.
Содержание SDM3055A
Страница 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Страница 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 113: ......
Страница 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...
Страница 140: ...Remote Manual SDM3055 SDM3055A Digital Mulimeter RC06035 E01A 2014 SIGLENT TECHNOLOGIES CO LTD ...