SIGLENT Remote Manual
57 SDM3055/SDM3055A
2.9.3 [SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO{ON|1|OFF|0}
[SENSe:]CURRent:{AC|DC}:NULL:VALue:AUTO?
Enables or disables automatic null value selection for AC or DC current measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0 (OFF) or 1 (ON)
Using automatic selection of zero value to measure:
CURR:AC:NULL:VAL:AUTO ON
READ?
Typical Response
:
-4.67956379E-08,+2.33978190E-08
When automatic reference selection is ON, the first measurement made is used as
the null value for all subsequent measurements.
[SENSe:]CURRent:{AC|DC}:NULL:VALue will be set to this value. Automatic null
value selection will be disabled.
When automatic null value selection is disabled (OFF), the null value is specified by
this command: [SENSe:] CURRent:{AC|DC}:NULL:VALue.
The instrument enables automatic null value selection when the null function is
enabled ([SENSe:]CURRent:{AC|DC}:NULL:STATe ON).
This parameter is set to its default value after a Factory Reset (*RST), Instrument
Preset (SYSTem:PRESet), or CONFigure function.
Содержание SDM3055A
Страница 1: ...User Manual SDM3055 Digital Multimeter UM06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 101: ...SIGLENT SDM3055 Digital Multimeter 91 Diagram 3 18 Result 2 ...
Страница 112: ...Quick Strat SDM3055 Digital Multimeter QS06035 E02A 2014 SIGLENT TECHNOLOGIES CO LTD ...
Страница 113: ......
Страница 123: ...SIGLENT 2 SDM3055 Quick Start Appearance and Size Front View Side View 260 mm 105 mm 282 mm ...
Страница 140: ...Remote Manual SDM3055 SDM3055A Digital Mulimeter RC06035 E01A 2014 SIGLENT TECHNOLOGIES CO LTD ...