Remote Manual
SIGLENT
SDM3055/SDM3055A 22
2.3.4 CALCulate:TRANsform:HISTogram:POINts{<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:POINts?[{MIN|MAX|DEF}]
Sets the number of bins between the lower and upper range values for the histogram. Two
additional bins always exist:one for measurements below the lower range and one for
measurements above the upper range.
Parameter
Typical
Return
{10|20|40|100|200|400|MIN|MAX|DEF},default
100
+100
See Example.
You can specify the lower and upper range values using
CALCulate:TRANsform:HISTogram:RANGe: {LOWer|UPPer}. Lower and upper
range values are computed automatically if
CALCulate:TRANsform:HISTogram:RANGe:AUTO is ON.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument
Preset (SYSTem:PRESet).