Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Force Transmit Procedure
This diagnostic software routine allows the user to “force” a transmitting condition that can be use to
search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The
routine forces the flowmeter to generate constant transmit bursts while reporting current receive signal
strength for the user. To initiate the
Force Transmit
function, refer to the example that follows.
1. On a PC, after the detection mode is selected, press <ENT>. If using the Magnetic Wand, the
detection mode is selected, touch <ENTER> sensor. While the flowmeter is going through the
drives (see menu screen below), press the <ALT> and <ENTER> keys simultaneously.
To complete the Install process disregard the following paragraphs explaining the Force Trans-
mit and Force Frequency diagnostic software routines. Proceed to the Install process comple-
tion steps immediately following the Force Transmit procedures.
NOTE: If the transducers have been installed successfully but the Estimated Vs (sonic veloc-
ity) has been changed, the Pick/Install Xdcr menu cell [Install Complete?] option list
will also display the [Transfer Install] function selection. The Transfer Install function
allows the transducers to be optimally positioned for a different fluid, without the need
for a new Initial Makeup procedure (see the Water Calibration addendum in this manual
for details).
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Direct
Spacing Offset
Minimum
Number Index
4
Spacing Method
Track 1012TP
Ltn Value (in)
0.217
Install Complete?
Install
Empty Pipe Set
Channel Not Setup
Zero Flow Adjust
Channel Not Setup
Siemens 2 Channel [1] SITE1
Drive 0
Pick/Install Xdcr
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Direct
Spacing Offset
Minimum
Number Index
4
Spacing Method
Track 1012TP
Ltn Value (in)
0.217
Install Complete?
No
Empty Pipe Set
Channel Not Setup
Zero Flow Adjust
Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Содержание SITRANS FUS1020
Страница 3: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......
Страница 219: ......
Страница 220: ......
Страница 221: ...21614 C ...
Страница 222: ...21614 C ...
Страница 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Страница 224: ...1011FTP 8 21614 C ...
Страница 225: ...1011FTNF 8 21614 C ...
Страница 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Страница 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Страница 228: ......