Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
307
Table A- 100 Path 5 diagnostics
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
3600
Float / 4
RxGain up path
5
Rx gain of upstream
signal
-
-
Read
only
3602
Float / 4
RxGain down
path 5
Rx gain of downstream
signal
-
-
Read
only
3604
Float / 4
SNR upstream
path 5
Signal to noise ration
of upstream signal
-
-
Read
only
3606
Float / 4
SNR down-
stream path 5
Signal to noise ration
of downstream signal
-
-
Read
only
3608
Float / 4
Sound velocity
pah 5
Sound velocity
-
-
Read
only
3610
Float / 4
Flow velocity
path 5
Flow velocity
-
-
Read
only
3612
Float / 4
Delta time path
5
Delta time
-
-
Read
only
3614
Float / 4
Travel time up
path 5
Travel time of up-
stream signal
-
-
Read
only
3616
Float / 4
Travel time
down path 5
Travel time of down-
stream signal
-
-
Read
only
3618
Float / 4
Correlation
factor path 5
Correlation factor
-
-
Read
only
3624
Unsigned / 2 Accepted burst
percentage
path 5
% of bursts accepted
(path 5).
Based on various di-
agnostic input (i.e.
correlation strength,
gain level, SNR, etc.)
the meter may reject
specific up/down re-
ceived sets or bursts.
The percentage of
accepted bursts is one
measure of the appli-
cation’s health.
-
-
Read
only
3625
Float / 4
Actual sensor
frequency path
5
Sensor frequency
-
-
Read
only
3629
Float / 4
Upstream peak
amplitude path
5
Upstream signal peak
amplitude
-
-
Read
only
3627
Float / 4
Downstream
peak amplitude
path 5
Downstream signal
peak amplitude
-
-
Read
only
3631
Float / 4
Fluid refraction
angle path 5
Calculated fluid refrac-
tion angle on path 5
(clamp-on only).
-
-
Read
only