Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
162
Operating Instructions, 08/2017, A5E37367739-AC
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
1312
Float / 4
Sensor crystal
projection
Length of the
crystal projec-
tion onto the
emitting surface
of the sensor.
0.03409
[m]
(8556)
0.0 - 3.0
Read /
write
1314
Float / 4
Sensor phase
velocity
The velocity at
which the wave
fronts traverse
the interface
between the
wedge and the
pipe.
3695.5
[m/s]
(7648)
2000.0 - 10000.0
Read /
write
1316
Float / 4
Sensor inactive
wedge
The part of the
wedge that is
not part of the
sound path.
0.00076
[m]
(8556)
0.0 - 0.05
Read /
write
1318
Float / 4
Sensor fixed
time
Fixed time in
sensor wedge.
10.61E-6
[s]
2.0E-6 - 200.0E-6
Read /
write
1320
Float / 4
Mounting hole
offset
Distance be-
tween spacer
hole and sensor
front face.
0.00953
[m]
(8556)
0.0 - 0.05
Read /
write
1322
Unsigned / 2 Pipe travel time Designated
multiplier on the
crystal projec-
tion value used
in formula for
calculating pipe
travel.
0
0 - 200
Read /
write
1711
Float / 4
Temperature
compensation
factor
Sensor tempera-
ture compensa-
tion slope factor.
0.00063761
[1/K]
0.0001 - 0.0009
Read /
write
1713
Unsigned / 2 Reference
temperature
Sensor refer-
ence (nominal)
temperature.
0
•
0: 21 °C
•
1: 60 °C
•
2: 93 °C
Read /
write