Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
159
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
1223
Unsigned / 2 Downstream
pipe geometry
Compensation for
downstream pipe ge-
ometry.
0
•
0: Straight run
•
1: Single elbow
•
2: Double elbow (in-
plane)
•
3: Double elbow (out-
of-plane)
•
4: Reducer
•
5: Expansion
Read /
write
1224
Unsigned / 2 Downstream
distance
The distance to the
downstream flow dis-
turbance in multiples of
inner pipe diameter.
0
0 - 40
Read /
write
Process medium characteristics
Table A- 17 Process medium characteristics
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
1825
Float / 4
Expected sound
velocity
Expected sound veloci-
ty in the medium at
medium temperature.
1500.0
[m/s]
(7648)
100.0 - 2500.0
Read /
write
1701
Float / 4
Fixed medium
temperature
Fixed medium temper-
ature value used for
compensation.
20
[°C]
(7700)
-273 - 400
Read /
write
1703
Float / 4
Fixed process
pressure
Fixed process pressure
value used for com-
pensation.
700000.0
[Pa]
(8015)
0.0 - 1E+8
Read /
write
1705
Float / 4
Fixed process
viscosity
Fixed process viscosity
value used for com-
pensation.
0.000001
[m
2
/s]
(7524)
0.0 - 5000.0
Read /
write
1707
Float / 4
Fixed process
density
Fixed process density
used for compensation.
1000.0
[kg/m
3
]
(7600)
0.0 - 2000.0
Read /
write