Parameters
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
83
3.1.5.1
Flow Direction
Main flow direction related to direction
arrow on measuring tube (= forwards,
positive flow values)
+Direction
•
+Direction,
•
-Direction
3.1.5.2
Meas.Direction Output of suppression of reverse flow
Affects all outputs, totalizers and display
forw.+rev.
•
forward only,
•
forw.+rev.
3.1.5.3
Hysteresis
Hysteresis of detection of flow direction in
% of full scale value
0.2 %
0 to 20 %
3.1.6
Low flow cut
Threshold for low flow cut in % of full
scale value 1, 2 or absolute value in flow
unit 3
1 %
0 to 20 %
3.1.7
Noise filter
3.1.7.1
Fltr.Time Con.
Time constant τ for measured value
attenuation; after a jump in
the measuring variable, the output
measured value reaches about 99% of
the new setpoint after 5*τ
Does not affect totalizers.
3.00 s
0.0 ... 200.0 s
3.1.7.2
Line Synchron.
Auto mode
not available
3.1.7.3
Noise Suppress
Suppress Time
Suppress Limit
Values outside a tolerance range are
evaluated for the duration of the blanking
time with high attenuation.
Time of effect of increased attenuation (0
s = blanking off)
Tolerance range = floating a
tolerance value
0 s
0 %
0 ... 100
0 ... 100
3.1.8
Slurry mode
This is used when signals are severely
impaired, e.g. for media with a high solids
content or air locks
Off
•
On
•
Off
3.2
Total. Forw.
Totalizer forwards
3.2.1
Set forward
Reset totalizer to zero and stop or start
Cancel 1,2
Totalize 3
•
Cancel 1,2,
•
Reset+Stop,
•
Reset+Start 1,2,
•
Totalize 3
3.2.2
Engr. Unit
Engineering unit of volume
or
Engineering unit of mass When you select
a unit of mass, entry of the actual medium
density is mandatory (menu 3.1.4)
DN 2...12: l
DN >12: m
3
•
l, hl, m
3
, Ml, ft
3
, Gal,
MGal, ImpGal,
MImpGal
•
kg, t, g, lb
3.2.3
Alarm Limit
+1000000 l
-108 ... +108
3.3
Total. rev.
Totalizer reverse
3.3.1
Set reverse
Reset totalizer to zero and stop or start
Cancel 1,2
Totalize 3
•
Cancel 1,2,
•
Reset+Stop,
•
Reset+Start 1,2,
•
Totalize 3
Содержание SITRANS F M TRANSMAG 2
Страница 1: ...Electromagnetic Flowmeters SITRANS F M TRANSMAG 2 Operating Instructions 11 2009 SITRANS F ...
Страница 2: ......
Страница 10: ...Safety notes 2 2 Laws and directives SITRANS F M TRANSMAG 2 8 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 14: ...Description 3 4 BUS Communication SITRANS F M TRANSMAG 2 12 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 40: ...Commissioning 6 4 Operating examples SITRANS F M TRANSMAG 2 38 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 92: ...Parameters SITRANS F M TRANSMAG 2 90 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 108: ...Profibus Communication B 9 Literature list SITRANS F M TRANSMAG 2 106 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 110: ...Appendix C 2 Accessories and spare parts SITRANS F M TRANSMAG 2 108 Operating Instructions 11 2009 A5E00102775 07 ...