Troubleshooting
9.3 Error messages
SITRANS F M TRANSMAG 2
68
Operating Instructions, 11/2009, A5E00102775-07
9.3
Error messages
Errors are indicated by a flashing "D" (Device error) or "F" (Measurement error) in the top
right hand corner of the local display.
Table 9- 2
Error messages
Error
Diagnosis
Remedy
D Flashing "D" in the top
right hand corner of the
display
General device error (see also section
5.3.2)
Menu 2.1 Call device status
D Device status (menu 2.1)
"Memory failure"
Internal memory defective
Replace complete electronics
HART interface not available in the
transmitter
D Device status (menu 2.1)
"COM Module failure"
PROFIBUS interface not available in
the transmitter
Replace complete electronics
D Device status (menu 2.1)
"Software failure"
General software error
Do a restart (menu 6.4);
If error persists: have firmware reloaded (by
service personnel), alternatively: replace
complete electronics
Magnetic field current fuse defective
Change magnetic field current fuse (see section
6.3.1)
Magnetic current cable defective (at
terminals 5 and 6 in remote design)
Check cable and change if necessary
Magnet coil defective
Check on remote design:
Switch off device, disconnect magnetic field
current cable, measure resistance at terminals 5
and 6 on sensor: setpoint: 8 ... 20 Ω.
Otherwise: change sensor
Reference coil improperly connected
(terminals 55, 66)
Check cable and replace if necessary
D Device status (menu 2.1)
"Sensor failure"
Electronics defective
Change complete electronics
D Device status (menu 2.1)
"Flow measurement
disturbed"
No measurement possible (group
message)
Change complete electronics
"SmartPLUG read failure" Connected sensor without SmartPLUG
or defective SmartPLUG
Check parameterization (range, unit etc.)
A correct measurement is still possible when the
calibration values CFH and ZPH match the
values on the rating plate of the sensor.
Check whether this is the case and correct if
necessary under the menus 6.7.2 and 6.7.3
F Flashing "F" in the top
right hand corner of the
display
General incorrect measurements,
process errors or incorrect
parameterization.
Menu 2.1 Call device status
F Device status (menu 2.1)
"Measurement module
failure"
Measuring range overflow
Reduce flow quantity or increase upper range
value in menu 3.1.2
F Device status (menu 2.1)
"Measurement module"
Measuring electronics defective
Measurement with reduced accuracy still
possible; replace complete electronics
Содержание SITRANS F M TRANSMAG 2
Страница 1: ...Electromagnetic Flowmeters SITRANS F M TRANSMAG 2 Operating Instructions 11 2009 SITRANS F ...
Страница 2: ......
Страница 10: ...Safety notes 2 2 Laws and directives SITRANS F M TRANSMAG 2 8 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 14: ...Description 3 4 BUS Communication SITRANS F M TRANSMAG 2 12 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 40: ...Commissioning 6 4 Operating examples SITRANS F M TRANSMAG 2 38 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 92: ...Parameters SITRANS F M TRANSMAG 2 90 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 108: ...Profibus Communication B 9 Literature list SITRANS F M TRANSMAG 2 106 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 110: ...Appendix C 2 Accessories and spare parts SITRANS F M TRANSMAG 2 108 Operating Instructions 11 2009 A5E00102775 07 ...