Functions
7.5 Function Group Device Outputs
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
55
scaled to the 0 to "Split value" range. If the flow value exceeds this subdomain, the output
current is scaled to the 0 to URV range.
Note
The relay contact is open in the "no function" setting.
● Signal type, menu item 4.3.2
You can set the function of the relay as follows in this menu item:
1.
When the relay need to be closed in the event of an alarm:
Flow in forward direction
Limit value message
2.
When the relay need to be open in the event of an alarm:
Flow in forward direction
Limit value message
Table 7- 3
Relay status
Setting
No alarm
Reverse flow
No limit signalling
Alarm message
Forward flow
Limit signalling
1. Contact closes
2. Contact opens
①
Electronic fuse
②
External supply
Figure 7-8
Digital output 2
Содержание SITRANS F M TRANSMAG 2
Страница 1: ...Electromagnetic Flowmeters SITRANS F M TRANSMAG 2 Operating Instructions 11 2009 SITRANS F ...
Страница 2: ......
Страница 10: ...Safety notes 2 2 Laws and directives SITRANS F M TRANSMAG 2 8 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 14: ...Description 3 4 BUS Communication SITRANS F M TRANSMAG 2 12 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 40: ...Commissioning 6 4 Operating examples SITRANS F M TRANSMAG 2 38 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 92: ...Parameters SITRANS F M TRANSMAG 2 90 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 108: ...Profibus Communication B 9 Literature list SITRANS F M TRANSMAG 2 106 Operating Instructions 11 2009 A5E00102775 07 ...
Страница 110: ...Appendix C 2 Accessories and spare parts SITRANS F M TRANSMAG 2 108 Operating Instructions 11 2009 A5E00102775 07 ...