Device
7.2 System
SIMATIC MV500
Operating Manual, 05/2019, C79000-G8976-C495-02
145
Version
Note
Licenses transferred to the reader with Automation License Manager
Licenses transferred to the reader with Automation License Manager (ALM) are not affected
by any of the possible management actions. The licenses are retained in all cases and do
not have to be saved beforehand.
Firmware update
You can run a firmware update with this function. To do so, select the desired update file
(.sfw) and then click "Start update".
Note
DISA bit must not be set during the firmware update
You must make sure that the DISA bit is not set during the firmware update. This prevents
an automation system from switching the reader to RUN.
For more detailed information, refer to the HTML page that opens.
Note
Prior to the firmware update: Saving settings and programs
All settings and programs are lost during the firmware update. Make sure that you save all
settings and programs before running an update.
Following a prompt for confirmation, the reader restarts in the update mode. The reader
retains the IP address that it had when the function was activated. If applicable, the "DHCP"
IP mode is disabled for the duration of the update.
For more detailed information, refer to the HTML page that opens.