Siemens-Elema AB
Troubleshooting
Servo
…
E382 E407E 061 01 01 02
6 - 8
6
16
EXP_FLOW_MTR_NODE_DISCON
1.
Replace PC1784 Expiratory channel.
17
EXP_FLOW_MTR_NODEID_CONFLICT
N/A
18
EXT_COM_DISCON
N/A
19
EXT_COM_NODEID_CONFLICT
N/A
22
BUZZER_SILENCER
1.
Replace PC1772 Monitoring.
24
BACKUP_CAP_ERR_MON
1.
Replace PC1772 Monitoring
25
TECH_MON_DEVICE_COMM_ERR #
Depending on #.
If repeated, replace parts as follows:
0:
PC 1770 Main back-plane.
1: PC
1772
Monitoring.
2:
PC 1775 Plug-and-play back plane.
3:
PC 1778 Standard connectors & DC/DC.
4:
PC 1781 Inspiratory pressure transducer.
5:
PC 1781 Expiratory pressure transducer.
6: O
2
cell or O
2
cell cable or PC 1772
Monitoring.
32: PC 1784 Expiratory channel or PC 1772
Monitoring.
33: Gas module Air or PC 1772 Monitoring.
34: Gas module O
2
or PC 1772 Monitoring.
35: PC1772
Monitoring.
256: PC1771 Control.
512: PC 1777 Panel.
784: N/A
1024: PC 1784 Expiratory channel.
1025: PC 1786 Expiratory channel cassette.
26
TECH_ERR_INCONSISTENT_MAINS_
FAILURE
N/A
27
TECH_ERR_BUZZER_FAILURE
1.
Replace PC1772 Monitoring.
Note: If the Patient Unit front cover is removed,
this error may be activated.
Содержание Servo i
Страница 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Страница 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Страница 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Страница 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Страница 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Страница 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Страница 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...