E382 E407E 061 01 01 02
Siemens-Elema AB
Servo
…
Troubleshooting
6 - 7
6
Technical error codes
The table below shows recommended actions in case
of Technical error alarms.
Error
code
Error message / Possible cause
Recommended action
Monitoring
1
POWER_FAILURE_MINUS_12_VOLTS_
TOO_LOW
1.
Check status of external battery (if
connected).
2.
Replace PC1778 Standard connectors &
DC/DC.
2
POWER_FAILURE_MINUS_12_VOLTS_
TOO_HIGH
1.
Replace PC1778 Standard connectors &
DC/DC.
3
POWER_FAILURE_12_VOLTS_TOO_LOW
1.
Check status of external battery (if
connected).
2.
Replace PC1778 Standard connectors &
DC/DC.
4
POWER_FAILURE_12_VOLTS_TOO_HIGH
1.
Replace PC1778 Standard connectors &
DC/DC.
5
POWER_FAILURE_24_VOLTS_TOO_LOW
1.
Replace PC1778 Standard connectors &
DC/DC.
2.
Replace gas modules.
6
POWER_FAILURE_24_VOLTS_TOO_HIGH
1.
Replace PC1778 Standard connectors &
DC/DC.
7
INSP_VALVE_RANGE_ERR
1.
Replace PC1771 Control.
2.
Replace PC1772 Monitoring.
3.
Replace gas modules.
8
TECH_ERR_INSP_PAUSE_HOLD_TIME_
EXCEEDED
1.
Replace PC1771 Control.
9
TECH_ERR_EXP_PAUSE_HOLD_TIME_
EXCEEDED
1.
Replace PC1771 Control.
10
VALVES_DISABLED
1.
Replace PC1784 Expiratory channel.
2.
Replace PC1771 Control.
3.
Replace PC1778 Standard connectors &
DC/DC.
11
SAFETY_VALVE_OPEN
1.
Check inspiratory channel
2.
Replace safety valve pull magnet.
3.
Replace PC1784 Expiratory channel.
12
BRE_NODE_DISCON
1.
Replace PC1771 Control.
13
BRE_NODEID_CONFLICT
N/A
14
PANEL_NODE_DISCON
1.
Check control cable
2.
Replace PC1777 Panel.
15
PANEL_NODEID_CONFLICT
N/A
Some of the Error codes are intended only for R&D,
not for field service. If so, the text ”N/A” is stated in
the ”Recommended action”-column.
Содержание Servo i
Страница 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Страница 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Страница 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Страница 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Страница 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Страница 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Страница 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...