Siemens-Elema AB
Service procedures
Servo
…
E382 E407E 061 01 01 02
5 - 8
5
Menu key
The ”Menu” function is a useful tool for the hospital
Biomeds as well as for the Siemens CSE (Customer
Service Engineer). There are no locked functions
within the Menu, all users have access to all
functions. However, the ”Biomed” submenu is
intended only for Technical Personnel.
Menu flowchart
Pressing the fixed key ”Menu” will open the ”Menu”
window on the User Interface. Possible selections in
this window, as described in the flowchart, are:
• Alarm
• Complience compensation
• Copy data
• Biomed
• Panel lock
• Change patient category (available only in ventilation
mode).
Biomed
Of the possible selections in the ”Menu” window,
the ”Biomed” menu will be of main interest for the
hospital Biomed and for the Siemens CSE.
Service
Event Log:
Displays log lists. Useful during troubleshooting.
Service Log:
Displays log lists. Useful during troubleshooting.
Report PM:
This button must be pressed when a Preventive
Maintenance has been performed. This will reset the
timer that indicates operating hours until next
Preventive Maintenance.
Panel
lock
General
Units
Adult alarm limits
Infant alarm limits
Displayed values
Displays the default user
configuration.
Locks all user input functions on the control unit.
Press "Main screen" fixed key to unlock.
View configuration
Adult alarm limits
Infant alarm limits
Displayed values
User default configuration setup.
Date and time setup.
Edit configuration
Set date and clock
Biomed
Service
Event log
Alarms
Ventilator settings
Apnea periods
Immediate functions
Displays Event logs.
Possible selections:
•
Latest logs
•
Selected time interval.
Service log
Report PM
Technical alarms
Test results
Preventive Maintenance
Service report history
Displays Service logs.
Possible selections:
•
Latest logs
•
Selected time interval.
Enters Date for a performed Preventive
Maintenance into the Service Log.
Alarm
Profile
Compliance
compensation
Copy
data
History
Alarm profile setup.
Possible selections:
•
Pressure alarm (Upper)
•
Minute Volume alarm (Lower and Upper)
•
Respiratory Rate alarm (Lower and Upper)
•
End Expiratory Pressure alarm (Lower)
•
Alarm sound level (20_100%).
Copy data to PC Card.
The following data is copied:
•
Event log
•
Trends
•
Pre-use check log.
Displays a date and time stamped Alarm history log.
Silence or pre-silence alarms for two minutes.
Same function as the fixed key .
The calculation of the circuit Compliance value is
performed during Pre-use check.
The calculated value is displayed in the window. To
activate or to deactivate select the appropriate soft key.
Menu
Change pat.
category
Available only in Servo-i Universal. Switches between
Adult and Infant patient category (only in running mode).
Содержание Servo i
Страница 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Страница 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Страница 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Страница 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Страница 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Страница 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Страница 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...