25
Siemens AG Healthcare Sector
11/12
Technical data
Ambient conditions
Control room
Temperature
15–30 °C (59–86 °F)
Relative air humidity
40 to 60%
Absolute air humidity
<11.0 g/kg
Electronics room
Temperature
15–30 °C (59–86 °F)
Relative air humidity
40 to 80%
Absolute air humidity
<11.0 g/kg
Examination room
Temperature
18–22 °C (64–72 °F)
Relative air humidity
40 to 60%
Absolute air humidity
<11.0 g/kg