Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
22
Tim table
Max. patient weight for vertical and horizontal table
movement
250 kg (550 lbs)
Max. scan range
140 cm; optional 205 cm
1
Vertical table movement Range
52–102 cm
2
+13 mm
3
Speed
6 cm/s
Horizontal table move-
ment
Max. range
275 cm
Max. speed
20 cm/s
Positioning accuracy
±0.5 mm
Continuous table movement during scan capable
1. With Tim Whole Body Suite option
2. Including Heightening Kit, if necessary
3. Depending on the floor conditions
Tim Dockable Table
Max. patient weight for vertical and horizontal table
movement
250 kg (550 lbs)
Max. scan range
140 cm; optional 205 cm
1
Vertical table movement Range
56–106 cm
2
+13 mm
3
Speed
6 cm/s
Horizontal table move-
ment
Max. range
275 cm
Max. speed
20 cm/s
Positioning accuracy
±0.5 mm
Continuous table movement during scan capable
1. With Tim Whole Body Suite option
2. Including Heightening Kit, if necessary
3. Depending on the floor conditions
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...