Safety information for the system owner
0.0
A.1-23
0.0
syngo MR D13
0.0
A.1
Refilling helium
A.1
The magnet is filled with liquid helium as a coolant. Following
installation, it is adjusted to the desired operating field strength.
A.1
During normal operation, the magnet does not lose helium.
Under special conditions - power failure, malfunctions of the
cold head and maintenance activities - liquid helium must be
refilled by Siemens Service.
A.1
When filling the magnet with helium, perform the necessary
tasks carefully and accurately, observing all regulations. Wear
protective clothing to prevent frostbite.
A.1
Helium-related risks
A.1
Liquid helium presents the following properties that, among
other things, may result in hazardous conditions when not han-
dled professionally:
A.1
■
Extremely cold: causes frostbites when it comes in touch
with skin
■
Oxygen in ambient air is displaced during boil-off: risk of
asphyxiation
If the helium fill level is too low, the alarm box or the
syngo
Acquisition Workplace will signal this accordingly.
A.1
◆
In case of alarm, notify Siemens Service and/or ensure refill-
ing only through trained and experienced personnel.
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...