11
Siemens AG Healthcare Sector
11/12
Technical data
Multi-Nuclear Option
1
•
Multi-Nuclear Imaging and Spectroscopy for the following nuclei:
3He, 7Li, 13C, 17O, 19F, 23Na, 31P, 129Xe
•
Decoupling nuclei: 13C, 31P
•
Hyperpolarization nuclei: 3He, 13C, 129Xe
Frequency band
3He
7Li
13C
17O
Center frequency [MHz]
93.8524
47.8799
30.9806
16.7012
Usable transmitter band-
width [kHz]
±50
±50
±50
±50
Usable receiver band-
width [kHz]
±190
±100
±65
±50
19F
23Na
31P
129Xe
Center frequency [MHz]
115.9040
32.5885
49.8719
34.0816
Usable transmitter band-
width [kHz]
±50
±50
±50
±50
Usable receiver band-
width [kHz]
±235
±65
±100
±70
1. Optional
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...