Siemens AG Healthcare Sector
11/12
Technical data
28
Power consumption
General classifications
System off
1
1. All data incl. cold head compressor, without cooling
5.0 kW
Stand-by
5.7 kW
Ready for measurement
13.9 kW
Typical measurement
20.7 kW
Protection class
I
Components used
•
Patient table
•
Local coils
•
Body coil
•
ECG/Pulse module
Type B applied part
Type B or BF applied part
Type B applied part
Type BF applied part
IP protection class according to
IEC 60529
IP X0
Explosion protection
The MR system is not intended for opera-
tion in areas prone to explosion (e.g.,
highly flammable mixtures of anaesthesia
gases with air or oxygen or nitrous oxide)
Operating mode
Continuous operation with short-term load
Disinfection receptors/system compo-
nents
Disinfectants without alcohol, ether
Degree of safety in the presence of a
flammable anesthetics mixture with air
or with oxygen or with nitrous oxide
No AP or APG category equipment
Mains operated equipment with addi-
tional power sources
None
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...