5
Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
Spatial distributions
This section includes the following spatial distributions diagrams:
•
Static magnetic field B
0
•
Spatial gradient of B
0
•
Product of the static magnetic field B
0
and the spatial gradient of B
0
Static magnetic field B
0
The figures show lines of the same magnetic flux density in milliTesla.
The following graphics show the calculated magnetic field in air. Magnetic materials in the
vincinity of the magnet (i.e. iron beams or room shielding) may influence the form of the
stray field.
The plots represent three orthogonal planes through the isocenter to illustrate maximum
spatial extent of iso-magnetic contours.
Each plot contains the iso-magnetic contours with values of 0.5 mT, 1 mT, 3 mT, 5 mT,
10 mT, 20 mT, 40 mT, and 200 mT as well as a distance scale and a superimposed outline
of the MR system.
The 0.5 mT line marks the exclusion zone of the static magnetic field (pace maker limit).
Note:
The magnetic field is rotationally symmetric about the z-axis and mirror symmetric
about the horizontal axis.
The multiple plots are shown to provide information about the accessible space in different
orientations.
This note also applies to the other magnetic field related plots.
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...