Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
12
Spatial gradient of the static magnetic field B
0
The rise of the magnetic field as a function of the distance to the magnet is expressed by
the spatial gradient of B
0
. The following figures show lines with the same gradient in T/m.
The magnetic attraction force on a magnetically saturated ferromagnetic object is propor-
tional to this quantity.
Please note: Sometimes this quantity is expressed in G/cm (1 T/m = 100 G/cm).
⊗
: At this location, the force on a magnetically saturated ferromagnetic object is greatest.
!
The iso-magnetic contour lines are accurate to a value better than 1%. There is very lit-
tle influence from the environment to the scanner. The graphic shows the cover in
nominal position. On a variety of scanners, the position can vary ±5 mm in the axial
direction. The maximum value therefore has a tolerance of ±10%, as small geometric
deviations can cause a significant change in the value.
−
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
1.2
1.4
1.6
1.8
0.2
0
0.2
0.4
0.6
0.8
1
3 T/m
3 T/m
1 T/m
0.5 T/m
0.3 T/m
0.1 T/m
5 T/m
5 T/m
7 T/m
10 T/m
11 T/m
0.1 T
/m
0.3 T/m
0.5 T/m
1 T/m
Z Axis (m)
V
e
rt
ical (m)
Patient table
Side view
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...