Operator Responsibilities for Maintenance
11
© Siemens, 2009
M7-000.664.01.04.02
Page 11 of 14
MAGNETOM
03.12
CS SD MR
Preventive maintenance
The purpose of preventive maintenance is to keep unforeseen failures to a minimum. This
satisfies the prerequisites for system compliance with the guaranteed, long-term charac-
teristics.
The effects of different operating conditions (full or partial load operation, temperature,
size of dust particles, humidity, gases, vapors) are checked and the condition of parts sub-
ject to wear is determined by recording and analyzing characteristic values. Preventive
measures must be adopted or repairs must be undertaken as appropriate.
The specified maintenance intervals correspond to the minimum requirements. Compli-
ance with stricter national legislation may occasionally be necessary.
Object or function
Reason
What is checked:
Interval
Cooling system
Preventive measure to
avoid overheating
Check for leaks and condensation
Check/replace water filter
Check water pressure and add water if
necessary
Annually
Air filters, fans
Preventive measure to
avoid pollution
Replace air filter
Check functionality of fans
Annually
Chiller (option)
Preventive measure to
avoid overheating
1
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...